! There is too much fragmentation at m/z<400
The basepeak cutoff is set to 1% because lower signals are mainly noise and they bear the danger for FP identifications.
The fragments indicative for the position are unreliable; the relative intensity of sn1/2 change from species to species.
Therefore, no position rules are defined.
All analytes are found, and there are no FP peaks identified.

04:	25:2_11:2	additional wrong FA composition reported	

36:2 in 02: no FA composition assigned
36:1 in 05: no FA composition assigned

All other FA compositions are correctly identified.


Biological samples:
2.3.pos5: For 36:4 and 38:4, there is an unrealistic compositions found; the plausible 20:4 combination is not detected because there is no partner chain found (20:4 fragment ion shows an unusually high intensity)
Plausible chains are detected for most species.
Sometimes seemingly apparent compositions are not found (38:5 in 2.3pos05).
Overall the identification is good, but negative ion mode is better suited for chain identification.