Standard:

Only 17:1 with correct retention time in (2) detected. 
Rules were developed based in the on 17:0, and on the other 3 spectra where the MS1 peaks
cannot be identified.

Not detected:

01:	16:0	not detected - there is noise present only
	17:1	not detected - there is noise present only
	18:0	not detected - there is noise present only
	18:1	not detected - MS1 peak seems to be too strong (but spectrum is present)
	

02:	16:0	not detected - there is noise present only
	18:0	not detected - there is noise present only
	18:1	not detected - covered by isotope of another species

03:	16:0	not detected - there is noise present only (but spectrum is present)
	17:1	not detected - there is noise present only (but spectrum present)
	18:0	not detected - covered by isotope of another species
	18:1	not detected - there is noise present only


Biological:
There are no MS/MS of LPS detectable. 
	